ICD / Pacemaker ERI (Elective Replacement Indicator)

Battery-depletion warning: device still works but tachytherapy or pacing may be compromised — OR requires external defibrillation and pacing backup

ERI is a battery-depletion warning ~3 months before true end-of-life; the device still works but ICDs typically disable tachytherapy first and pacemakers may downgrade to a backup mode, so external defibrillation and transcutaneous pacing must be immediately available in the OR.

If you forget

ERI ≠ EOL — device still works; ERI fires ~3 months before EOL

Board trap
ICD flagged 'ERI' on the chart and you place a magnet expecting shock therapy to be inhibited — but the device may have already auto-disabled tachytherapy to save the battery. If VT/VF occurs, deliver external defibrillation immediately; do not wait for the device to fire.

Key numbers

  • ERI fires ~3 months before EOL (manufacturer-specific, ~2.5–2.8 V RRT threshold)
  • EMI is significant if the source is <15 cm from the CIED generator
  • Maintain intraop MAP ≥60–65 mmHg or SBP ≥90 mmHg
  • External pacing AND external defibrillation capability immediately available for every CIED case (apply pads when EMI or pacing-dependent)
Reviewed Sep 24, 2026 for clinical accuracy by BoardsBridge Editorial Process

Uri Rozen, MD — Licensed Physician, Israel

Source guidelines: AHA ACC Perioperative Cardiovascular 2024 · AHA Adult ALS 2025 · ACC AHA Valvular Heart Disease 2020

Re-review cadence: 12 months from reviewed date.

Drill this as a SOE stem →

Related on BoardsBridge

Card ID: icd-pacemaker-eri · Path: aba.cardiac.arrhythmias-and-devices.icd-pacemaker-eri

Educational reference only. Reviewed for clinical accuracy against current authoritative guidelines. Not a substitute for clinical judgment. See editorial process and the sitemap entry for this card.