ICD / Pacemaker ERI (Elective Replacement Indicator)
Battery-depletion warning: device still works but tachytherapy or pacing may be compromised — OR requires external defibrillation and pacing backup
ERI is a battery-depletion warning ~3 months before true end-of-life; the device still works but ICDs typically disable tachytherapy first and pacemakers may downgrade to a backup mode, so external defibrillation and transcutaneous pacing must be immediately available in the OR.
If you forget
ERI ≠ EOL — device still works; ERI fires ~3 months before EOL
Board trap
Key numbers
- ERI fires ~3 months before EOL (manufacturer-specific, ~2.5–2.8 V RRT threshold)
- EMI is significant if the source is <15 cm from the CIED generator
- Maintain intraop MAP ≥60–65 mmHg or SBP ≥90 mmHg
- External pacing AND external defibrillation capability immediately available for every CIED case (apply pads when EMI or pacing-dependent)
Uri Rozen, MD — Licensed Physician, Israel
Source guidelines: AHA ACC Perioperative Cardiovascular 2024 · AHA Adult ALS 2025 · ACC AHA Valvular Heart Disease 2020
Re-review cadence: 12 months from reviewed date.