IVC POCUS for Volume Status

Bedside inferior vena cava ultrasound to estimate right atrial pressure and fluid responsiveness in the perioperative patient

Bedside ultrasound of the inferior vena cava estimates right-sided filling pressures by correlating IVC diameter and respiratory collapsibility with central venous pressure and fluid responsiveness.

If you forget

IVC >2.1 cm + <50% collapse = elevated CVP (~10-15 mmHg) in spontaneous breathing

Board trap
A plethoric, non-collapsing IVC does NOT equal fluid responsiveness — it reflects elevated right-sided pressure; the patient may be in RV failure and need inotropes/vasopressors, not more crystalloid.

Key numbers

  • IVC >2.1 cm + <50% collapse → CVP ~15 mmHg (10-20)
  • IVC <2.1 cm + >50% collapse → CVP ~3 mmHg (0-5)
  • Mechanical ventilation: IVC >1.2 cm + <12% Δ diameter → elevated CVP
  • Measurement site: 1-2 cm caudal to the cavoatrial junction (avoid at the junction itself)
  • Collapsibility Index (CI) = (Dmax − Dmin) / Dmax × 100